Local Level Lab

Airborne measurements. Open methods.

Why calibration stopped

Both starts of one required setting failed the numerical cost-stability check. The phase therefore stopped under its predeclared rule; validation did not start.

This review reads eight records selected before inspecting their fit details: seven saved fits and one interrupted start. The failed pair is compared with the preceding noise example and the available alternate-processing fit. No observations were generated, no fits were rerun and no native calculations occurred.

Saved check Failed pair Required limit
Evaluations used 41 each At most 200
Fresh stationarity About 0.0000356 At most 0.0001
Largest fresh-check position RMS change About 0.225 mm At most 1 mm
Largest fresh-check position change About 0.482 mm At most 10 mm
Relative cost change About 0.000000827 At most 0.00000001

The cost change is about 83 times its frozen tolerance. The predicted positions pass their physical-error screens, but that does not override numerical qualification. Both failed starts report “no positive predicted decrease.” The preceding example passes all four selected starts; the available alternate processing case also passes. The interrupted flat-model start has no completed fit and is not treated as a numerical or physical model failure.

The reusable saved-record reader and reporting adapter passed seven software checks. The immutable evidence snapshot preserves the request, exact selected records, their hashes, reporting source and unchanged failed labels. Reproduction uses lll.prediction_phase_review.run and the frozen request in a separate working copy; the original one-review allowance is spent.

Selected records CSV · Review plan · Phase outcome · Calibration evidence.